Analytical Formula for the total scatter factor of the Small Radiation Field Size

J. Al Atawneh, S (2021) Analytical Formula for the total scatter factor of the Small Radiation Field Size. Journal of Research in Applied and Basic Medical Sciences, 7 (3). pp. 152-160.

[img] Text
7 Saed J A-10-164-1.pdf

Download (832kB)

Abstract

We present an analytical formula (AF) to calculate the total scatter factor of the MLC-shaped small radiation beams. Materials & Methods: Pinpoint, Semiflex, and Farmer ion chambers were used for MLC/Jaw-shaped fields extending from (10x10 cm2) down to (1x1 cm2). The total scatters factor, beam profiles, and penumbra was investigated for MLC/Jaw-shaped beams. Results: We found that the total scatters factor and the penumbra was clearly higher for MLC-shaped beams due to the MLC leakage, which becomes more and more significant for the small fields. The Analytical Formula was introduced as a least-square fit equation and verified against to measure data. The maximum deviation at small field size (< 4 cm2) was within 3.6% and 43.9% from both MLC and Jaw-shaped beams, respectively. Conclusion: Our analytical formula shows a very good agreement with measured data at a large field size (> 4 cm2), and had small deviation with MLC-shaped small field size, emphasizing that our analytical formula is valid and reasonably good for the total scatters factor calculation of the MLC-shaped beams, especially the ultra-small field size.

Item Type: Article
Uncontrolled Keywords: Small field dosimetry, Total Scatter Factor, Analytical Formula of Small Radiation Field Size
Subjects: R Medicine > R Medicine (General)
Depositing User: Unnamed user with email gholipour.s@umsu.ac.ir
Date Deposited: 21 Feb 2022 06:30
Last Modified: 21 Feb 2022 06:30
URI: http://eprints.umsu.ac.ir/id/eprint/6407

Actions (login required)

View Item View Item